General Specifications
|
Power Requirements
|
100 V ac to 240 V ac (-15 %, +10 %), 50 Hz to 60 Hz
|
Maximum Power Consumption
|
Controller: 110 VA
Intensifier: 250 VA
|
Operating Temperature Range
|
0 °C to 50 °C
|
Storage Temperature Range
|
-30 °C to 71 °C
|
Humidity
|
5 % to 95 %
|
Vibration
|
Meets MIL-T-28800D
|
Altitude (operation)
|
2,000 m
|
Weight
|
Controller: 12 kg (27 lb)
Intensifier: 14 kg (31 lb)
Accessory Kit with Handheld Control: 12 kg (27 lb)
Nitrogen Cylinder: 12 kg (26 lb)
|
Dimensions
|
Controller, Intensifier, and Accessory Kit: 32.5 cm Height x 30.5 cm Width x 30.5 cm Depth (13 in Height x 12 in Width x 12 in Depth)
Nitrogen Cylinder: 68.5 cm Height x 17.8 cm Diameter (27 in Height x 7 in Diameter)
|
Remote Communications Interfaces
|
RS-232, USB type B, IEEE-488.2, USB type A (PRINTER port only)
|
Pressure Range
|
3.5 kPa (0.5 psi) absolute to 70 MPa (10,000 psi)
|
Operating Medium
|
Dry nitrogen supply, or ambient air compressed by on-board pump
|
Warm-up Time
|
15 minutes fr om power on, after at least 60 minutes ambient temperature acclimation time within the working temperature range
|
Controller, intensifier, and nitrogen cylinder connections
Unit
|
Name
|
Type of Fittings [1]
|
Maximum Working
Pressure Limit
|
Controller
|
TEST Port
|
IQC70 female, located on CPS
|
70 MPa (10,000 psi)
|
SUPPLY Port
|
IQC70 female
|
Hose Purge
|
IQC70 male
|
Intensifier
|
SUPPLY Port
|
IQC70 female
|
70 MPa (10,000 psi)
|
OUTLET Port
|
IQC70 female
|
Nitrogen Cylinder
|
VALVE
|
CGA580 nipple
|
40 MPa (6,000 psi)
|
Regulator Outlet
|
Miniature QC female
|
20 MPa (3,000 psi)
|
Notes:
- ICQ70 is a Fluke Calibration interlocking quick connection nominally rated to 70 MPa (10,000 psi).
|
Accessory kit fittings
Name
|
Type of Fittings[1]
|
Maximum Working
Pressure Limit[2]
|
FITTING, 1/2 MALE PIPE THREAD (NPT)
|
IQC70 male
|
70 MPa (10,000 psi)
|
FITTING, 1/4 MALE PIPE THREAD (NPT)
|
FITTING, 1/8 MALE PIPE THREAD (NPT)
|
FITTING, 1/2 FEMALE PIPE THREAD (NPT)
|
FITTING, 1/4 FEMALE PIPE THREAD (NPT)
|
FITTING, 1/8 FEMALE PIPE THREAD (NPT)
|
FITTING, HF4 EQUIVALENT
|
FITTING, AN6 FEMALE
|
FITTING, AN4 FEMALE
|
FITTING, HM4 EQUIVALENT
|
FITTING, AN6 MALE
|
FITTING, AN4 MALE
|
FITTING,1/4 ID HOSE BARB
|
IQC70 male
|
700 kPa (100 psi)
|
FITTING, 1/8 ID HOSE BARB
|
IQC70 male
|
700 kPa (100 psi)
|
Notes:
- IQC70 is a Fluke Calibration interlocking quick connection nominally rated to 70 MPa (10,000 psi).
- Working pressures indicated are for adaptors only. Actual pressure capability may be limited by the integrity or maximum working pressure of the fitting to which the adaptor is connected.
|
Test hoses
Name
|
Type of Fittings [1]
|
Maximum Working
Pressure Lim it
|
Test Hose 3 m (10 ft)
|
IQC70 male at one end, and IQC70 female coupling at the other
|
70 MPa (10,000 psi)
|
Interconnect hose 1.8 m (6 ft)
|
IQC70 male at both ends
|
70 MPa (10,000 psi)
|
Supply hose 1.8 m (6 ft)
|
IQC70 male at one end, and miniature QC at the other
|
20 MPa (3,000 psi)[2]
|
Notes:
- IQC70 is a Fluke Calibration interlocking quick connection nominally rated to 70 MPa (10,000 psi).
- Hose limited to 3,000 psi due to miniature QC used to connect to the fitting on the regulator.
|
Measurement specifications
Barometer Module Range
|
70 kPa to 110 kPa absolute (10 psi to 16 psi)
|
Pressure Transducer Module Ranges [1]
|
A150k
|
3.5 kPa to 150 kPa absolute (0.5 psi to 21.5 psi)
|
G700k
|
50 kPa to 700 kPa gauge (7 psi to 100 psi)
|
G10M
|
700 kPa to 10 MPa gauge (100 psi to 1,500 psi)
|
G70M
|
10 MPa to 70 MPa gauge (1,500 psi to 10,000 psi)
|
Resolution
|
To greater of 1 ppm of reading or of 1ppm of 4 psi (0.00004 psi or 2.5 Pa)
|
Measurement uncertainty
Gauge, Absolute,
or Vacuum
|
Range
|
Uncertainty[1]
|
Gauge[2]
|
-98 kPa to +25 kPa
(-14.2 to +4 psi)
|
±25 Pa (0.004 psi,
0.1 in H2O at 20°C)
|
25 kPa to 70 MPa
(4 to 10,000 psi)
|
±0.1 % of reading
|
Absolute
|
3.5 to 130 kPa
(0.5 to 19 psi) absolute
|
±25 Pa (0.004 psi,
0.1 in H2O at 20°C)
|
130 kPa to 70 MPa
(19 to 10,000 psi) absolute
|
±0.1 % of reading
|
Vacuum[2]
|
0 kPa to 98 kPa
(0 to 29 inHg)
|
±25 Pa (0.007 inHg)
|
Notes:
[1] Measurement uncertainty defined as the maximum deviation from the indicated value of measured pressure for 1 year after alignment including hysteresis, linearity, repeatability, stability, temperature, humidity effects, and calibration reference standard measurement uncertainty. Expression of uncertainty uses a coverage factor of 2 and conforms with the recommendations of the ISO Guide to the Expression of Uncertainty in Measurement.
[2] Negative gauge and vacuum pressure values assume the ambient atmospheric pressure is approximately 101 kPa (14.7 psi) absolute.
|
Pressure control specifications
Control Precision
|
greater of ± 28 Pa (0.004 psi) or 0.025 % of reading
|
Control ranges
Pressure Source
|
Gauge, Absolute,
or Vacuum
|
Range
|
On-board Pump
|
Gauge[1]
|
-98 kPa (-14.2 psig) to
2 MPa (300 psig)
|
|
Absolute
|
3.5 kPa (0.5 psi) to
2 MPa (300 psi) absolute
|
|
Vacuum (negative gauge) [1]
|
0 to 98 kPa (29 inHg)
|
External Gas Supply
|
Gauge
|
28 kPa (4 psig) to
70 MPa (10,000 psig)
|
|
Absolute
|
130 kPa (19 psi) to
70 MPa (10,000 psi) absolute
|
[1] Negative gauge and vacuum pressure values assume the ambient atmospheric pressure is approximately 101 kPa (14.7 psi) absolute.
|
Typical control set times
Measure Mode
|
Range
|
Time (seconds) [5]
|
Gauge [7]
|
<28 kPa (4 psig)
|
90 [1] [6]
|
|
28 to 700 kPa (4 to 100 psig)
|
30 [2]
|
|
700 kPa (100 psig) to 7 MPa (1,000 psig)
|
30 [3]
|
|
7 MPa (1,000 psig) to 70 MPa (10,000 psig)
|
60 [4]
|
Absolute
|
<14 kPa (2 psi) absolute
|
180 [1] [6]
|
|
14 to 130 kPa (2 to 19 psi) absolute
|
90 [1] [6]
|
|
130 to 700 kPa (19 to 100 psi) absolute
|
30 [2]
|
|
700 kPa (100 psi) to 7 MPa (1,000 psi)
|
30 [3]
|
|
7 MPa (1,000 psi) to 70 MPa (10,000 psi)
|
60 [4]
|
Vacuum [7]
|
|
|
|
>88 kPa (26 inHg) vacuum
|
180 [1] [6]
|
|
0 to 88 kPa (0 to 26 inHg) vacuum
|
90 [1] [6]
|
[1] Assumes 1pp cc test volume, uses on-board pump control
[2] Assumes 100 cc test volume with external gas supply.
[3] Assumes 50 cc test volume with external gas supply.
[4] Assumes 25 cc test volume with external gas supply.
[5] Set times assume pressure steps of no greater than 10% of target pressure and reflect time to ready using default control limits in dynamic control mode.
[6] Times for sub-atmospheric set points assume that the temperature is below the boiling point of any liquid in the UUT and CPS for the target pressure.
[7] Negative gauge and vacuum pressure values assume the ambient atmospheric pressure is approximately 101 kPa (14.7 psi) absolute.
|
|